- Testing Equipment[6]
- Other Analysis Instruments[6]
- Other Measuring & Analysing Instruments[6]
- Other Chemical Equipment[2]
- Instrument Parts & Accessories[1]
- Other Measuring & Gauging Tools[6]
- Other Machinery[1]
- Other Electronic Measuring Instruments[3]
- Gas Analyzers[1]
- Drying Equipment[1]
- Carbon Analyzers[3]
- Moisture Meters[2]
- Level Measuring Instruments[1]
- Measuring & Analysing Instrument Stocks[1]
- Contact Person : Ms. Tong Jane
- Company Name : North Dalian Analytical Instrument Co., Ltd.
- Tel : 86-411-84754333-306
- Fax : 86-411-84753759
- Address : Liaoning,Dalian,No.4-1 Rongshen Industry Park, No.2 Xuezi St., Qixianling Rd., Hi-tech Zone
- Country/Region : China
- Zip : 116023
Related Product Searches:BF-28-2 Tester for Oxidation Stability of Steam Turbine Oils by Rotating Pressure Vessel,High Quality,rotating bomb, rotating bomb,BF-28-2
BF-28-2 Tester for Oxidation Stability of Steam Turbine Oils by Rotating Pressure Vessel
1. Performance and Characteristics
Single-chip control; 8-inch color LCD; touch screen operation; automatically charge and release the oxygen; automatically show the oxidation curve; automatically store and print the test result; oil bath; two bombs.
2. Technical Parameters
Working power | AC 220V±10% 50HZ |
Input power | ≤3000W |
Pressure range | 0~1500kpa |
Pressure measurement accuracy | ±2‰kpa |
Temp. range by oil bath | 100°C~160°C optional |
Temp. control accuracy | ±0.1°C |
Volume of sample | 50±0.5g |
3. Applicable Standards
BF-28-2 conforms to the standards of SH/T 0193, ASTM D2272.
4. Scope
This instrument utilizes an oxygen-pressure vessel to evaluate the oxidation stability of new and in-service turbine oils having the same composition (base stock and additives) in the presence of water and a copper catalyst coil at 150°C.
5. Summary of Test Method
The test oil, water, and copper catalyst coil, contained in a covered glass container, are placed in a vessel equipped with a pressure gage. The vessel is charged with oxygen to a gage pressure of 620kpa (90 psi, 6.2 bar), placed in a constant-temperature oil bath set at 150°C, and rotated axially at 100 rpm at an angel of 30°from the horizontal. The number of minutes required to reach a specific drop in gage pressure is the oxidation stability of the test sample.